Обзор

Область применения



The following table lists the measuring conditions for standard applications. The listed values for the measuring range and detection limit (DL) are only approximate values. The exact values at the respective measuring point depend on the totality of all influencing variables and can be determined by Siemens for the specific case. Please note that the values for the detection limit and the maximum measuring range refer to a path length of 1 m. Longer path lengths will improve the detection limit, but not linearly. due to limiting effects such as dust load. The maximum applicable measuring ranges can only be used if permitted by the process conditions such as dust load.

Standard application

Effective optical path length: 0.3 … 12 m

Dust load2): < 50 g/Nm3

Process gas temperature

Tmin  … Tmax

Process gas pressure pmin  … pmax

Min. measuring range

(with 1 m eff. opt. path length)

Max. measuring range

(also dependent on eff. opt. path length: see following column)

(Max. measuring range x path length)

(DL x path length)

Accuracy3)

                 

under standard conditions1) without cross-interferences of other gases

At 1 013 hPa with cross-interference of gas 2

 

Gas 1

Gas 2

Gas code

Appl. code

   

Gas 1

Gas 1

Gas 1

Gas 1

Gas 1

Gas 1

O2

 

A

B6)

600 … 1 200 °C

950 …1 050 hPa

0 … 15 vol%

0 … 100 vol%

240 vol%*m

0.3 vol%*m
At 600 °C

 

5 %

C

0 … 600 °C

950 …1 050 hPa

0 … 5 vol%

0 … 100 vol%

75 vol%*m

0.1 vol%*m

 

2 %4)

P

0 … 200 °C

950 … 5 000 hPa

0 … 5 vol%

0 … 100 vol%

75 vol%*m

0.1 vol%*m

 

2 %

NH3

 

C

A

0 … 150 °C

950 … 1 050 hPa

0 … 25 ppmv

0 … 500 ppmv

2 500 ppmv*m

0.5 ppmv*m

0.9 ppmv*m

At 15 vol% H2O, 55 °C

2 %

T

0 … 150 °C

950 … 1 050 hPa

0 … 25 ppmv

0 … 500 ppmv

2 500 ppmv*m

0.5 ppmv*m

0.9 ppmv*m

At 15 vol% H2O, 55 °C

2 %

E

250 … 350 °C

950 … 1 050 hPa

0 … 45 ppmv

0 … 500 ppmv

2 500 ppmv*m

0.9 ppmv*m

At 250 °C

1.4 ppmv*m

At 15 vol% H2O, 250 °C

2 %

F

300 … 400 °C

950 … 1 050 hPa

0 … 50 ppmv

0 … 500 ppmv

2 500 ppmv*m

1 ppmv*m

At 300 °C

1.5 ppmv*m

At 15 vol% H2O, 300 °C

2 %

NH3

H2O

D

A

0 … 150 °C

950 … 1 050 hPa

0 … 25 ppmv

0 … 100 ppmv

1 200 ppmv*m

0.5 ppmv*m

0.9 ppmv*m

At 15 vol% H2O, 55 °C

2 %

T

0 … 150 °C

950 … 1 050 hPa

0 … 25 ppmv

0 … 100 ppmv

1 200 ppmv*m

0.5 ppmv*m

0.9 ppmv*m

At 15 vol% H2O, 55 °C

2 %

E

250 … 350 °C

950 … 1 050 hPa

0 … 45 ppmv

0 … 100 ppmv

1 200 ppmv*m

0.9 ppmv*m

At 250 °C

1.4 ppmv*m

At 15 vol% H2O, 250 °C

2 %

F

300 … 400 °C

950 … 1 050 hPa

0 … 50 ppmv

0 … 100 ppmv

1 200 ppmv*m

1 ppmv*m

At 300 °C

1.5 ppmv*m

At 15 vol% H2O, 300 °C

2 %

HCl

 

E

A

0 … 150 °C

950 … 1 050 hPa

0 … 30 ppmv

0 … 6 000  ppmv

1 200 ppmv*m

0.6 ppmv*m

2.2 ppmv*m

At 15% H2O, 55 °C

5 %

T

120 … 210 °C

950 … 1 050 hPa

0 … 10 ppmv

0 … 60 ppmv

720 ppmv*m

     

H

150 … 250 °C

950 … 1 050 hPa

0 … 50 ppmv

0 … 6 000 ppmv

1 200 ppmv*m

1.0 ppmv*m

At 150 °C

3.1 ppmv*m

At 15 vol% H2O, 150 °C

5 %

HCl

H2O

F

A

0 … 150 °C

950 … 1 050 hPa

0 … 30 ppmv

0 … 100 ppmv

1 200 ppmv*m

0.6 ppmv*m

2.2 ppmv*m

At 15% H2O, 55 °C

5 %

T

120 … 210 °C

950 … 1 050 hPa

0 … 10 ppmv

0 … 60 ppmv

720 ppmv*m

     

H

150 … 250 °C

950 … 1 050 hPa

0 … 50 ppmv

0 … 100 ppmv

1 200 ppmv*m

1.0 ppmv*m

At 150 °C

3.1 ppmv*m

At 15 vol% H2O, 150 °C

5 %

HF

 

G

A

0 … 150 °C

950 … 1 050 hPa

0 … 5 ppmv

0 … 1 500 ppmv

200 ppmv*m

0.1 ppmv*m

0.6 ppmv*m

At 15 vol% H2O, 55 °C

5 %

H

150 … 250 °C

950 … 1 050 hPa

0 … 5 ppmv

0 … 1 500 ppmv

200 ppmv*m

0.11 ppmv*m

At 150 °C

0.6 ppmv*m

At 15 vol% H2O, 150 °C

5 %

HF

H2O

H

A

0 … 150 °C

950 … 1 050 hPa

0 … 5 ppmv

0 … 200 ppmv

200 ppmv*m

0.1 ppmv*m

0.6 ppmv*m

At 15 vol% H2O, 55 °C

5 %

H

150 … 250 °C

950 … 1 050 hPa

0 … 5 ppmv

0 … 200 ppmv

200 ppmv*m

0.11 ppmv*m

At 150 °C

0.6 ppmv*m

At 15 vol% H2O, 150 °C

5 %

CO

 

J

C

0 … 600 °C

950 … 1 050 hPa

0 … 1.5 vol%

0 … 100 vol%

40 vol%*m

300 ppmv*m

1 500 ppmv*m

at 50 vol% CO2, 20°C

2 %

CO

CO2

K

D

0 … 400 °C

800 …1 400 hPa

0 … 3.0 vol%

0 … 100 vol%

35 vol%*m

0.06 vol%*m

Заказные данные

Заказной № Вес, кг Описание Наличие на складе Заказать
7MB6121-.....-.... 14 LDS 6 ДИОДНОЛАЗЕРНЫЙ СПЕКТРОМЕТР, ЦЕНТРАЛЬНЫЙ МОДУЛЬ ------ Уточните конфигурацию

Аксессуары

Заказной № Вес, кг Описание Наличие на складе Заказать
A5E00534675 1.5 LDS 6, Calibration verification kit NH3, version 1 ------ Заказать
A5E00822928 0.11 Ldscom communication software ------
A5E00823339003 4.8 LDS 6, TUEV/MCERT Linearity verification kit NH3 , version 1, 2 cells ------ Заказать
A5E00823339004 5.8 LDS 6, TUEV/MCERT Linearity verification kit NH3/H2O, version 1, 3 cells ------ Заказать
A5E00823339005 4.8 LDS 6, TUEV/MCERT Linearity verification kit H2O , version 1, 2 cells ------ Заказать
A5E00823339008 4.8 LDS 6, TUEV/MCERT Linearity verification kit HCl , 2 cells ------ Заказать
A5E00823339009 4.8 LDS 6, TUEV/MCERT Linearity verification kit HCl/H2O, 3 cells ------ Заказать
A5E00823339013 4.8 LDS 6, TUEV/MCERT Linearity verification kit NH3 , version 2, 2 cells ------ Заказать
A5E00823339014 5.8 LDS 6, TUEV/MCERT Linearity verification kit NH3/H2O, version 2, 3 cells ------ Заказать
A5E00823339015 4.8 LDS 6, TUEV/MCERT Linearity verification kit H2O , (only for H2O analyzer) Version 2, 2 cells ------ Заказать
A5E00854188 3 LDS 6, External power supply for 3 channels ------ Заказать
A5E01075594 3.8 LDS 6, Calibration verification kit NH3, version 2 ------ Заказать
Неверный логин или пароль!
Оформить заказ
Очистить
Каталог (по буквам)
Корзина
почта: info@teslacom.ru
тел./факс: 8 (495) 229-56-79
OOO «ТеслаКом» тел./факс (495)229-56-79, info@teslacom.ru